The PC-10 manual dictates a specific sequence for creating a pipette 1.2.1. Step 1: Single Stage Pulling (For Preliminary Tapering) Set the Mode-selector knob to STEP 1 . Adjust the NO. 1 HEATER level to the desired value. Press the Start switch. Step 2: Two-Stage Pulling (For Finalizing Tips) For a two-stage pull (essential for finer tips): Set the mode to STEP 2 . Set the NO. 1 HEATER level.

| Problem | Likely Cause | Fix | |--------|-------------|-----| | Glass melts but does not pull | Magnet power too low | Increase magnet power | | Glass breaks into pieces | Too much heat or too fast pull | Lower heater current or reduce weight | | Tip too long & floppy | Insufficient heat | Increase heater current or timer | | Pipette bent at tip | Uneven heating | Balance left/right heater currents | | No heating | Blown heater coil | Replace coil (order from Narishige) |

This public link is valid for 7 days and shares a thread, including any personal information you added. This link or copies made by others cannot be deleted. If you share with third parties, their policies apply. Can’t copy the link right now. Try again later. narishige pc10 manual new

Comprehensive Guide and User Manual for the Narishige PC-10 Glass Pipette Puller

The updated Narishige PC-10 manual includes the following new features: The PC-10 manual dictates a specific sequence for

Regular maintenance guarantees long-term repeatability and preserves the calibration of your Narishige PC-10. Cleaning Protocols

Decrease the heater dial value or reduce the secondary pull weight. Pulling force is too high / Heat too high Remove a weight or lower the heater setting. Unequal pipette shapes Capillary was not centered in the coil 1 HEATER level to the desired value

Loosen clamps and re-center the capillary vertically and horizontally within the loop. Heater setting too low or coil is broken

: Digital LED interface indexing current levels up to a maximum value of 100.